Model Number EG27-I10
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  malfunction  
Event Description
The time of event is not during procedure.There was no report of patient harm.Angulation stuck.
 
Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the angulation down stuck in l position.Based on the result, we concluded that it was caused due to the excessive force applied on the angulation down.Based on the technical report ""hr-rpt-0587(angle)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO GASTROSCOPE - I10 SLIM
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key17849589
MDR Text Key324671544
Report Number9610877-2023-59498
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04961333186846
UDI-Public04961333186846
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K131902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG27-I10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2023
Date Manufacturer Received09/06/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-