The mayfield composite series skull clamp (a3059) was returned for evaluation: device history record (dhr) review: the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - investigation of the returned unit showed that unit received had up and down movement in the lock, and the teeth were grinding in the unlocked position.All worn components were replaced with new parts, and general cleaning and maintenance were performed.Root cause: complaint confirmed via inspection of the unit.The unit required replacement of worn components due to routine use and wear.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|