RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Volume Accuracy Problem (1675)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2023 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint by the biomedical (biomed) engineer on the v60 indicating that the tidal volume was showing all zeros.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.The biomedical engineer (biomed) informed the remote service engineer (rse) that the tidal volume was showing all zeros and also noticed that the average air was reading 4.0 slpm while troubleshooting.The rse provided the biomed with the latest service manual, recommended that the biomed should replace the flow sensor assembly, and provided the biomed with the replacement part number for repair.The biomed informed the rse that the flow sensor assembly would be replaced.Investigation is ongoing.
|
|
Manufacturer Narrative
|
Multiple attempts have been made on 02-oct-2023, 09-oct-2023, and 19-oct-2023 to try to obtain further information about this case, but no response was received from the customer.This file is closed and can be reopened if new information becomes available.
|
|