Related manufacturer reference numbers: 2017865-2023-47561.It was reported that the patient presented in clinic for follow-up.Upon chest x-ray examination, it was found that the right ventricular (rv) lead had dislodged.During rv lead repositioning procedure, it was found that the rv lead helix failed to be retracted, and the rv lead was unable to be disconnected from the implantable cardioverter defibrillator (icd).Both the icd and the rv lead were explanted and replaced.There were no patient consequences.
|