The user facility reported a 69-year-old male undergoing cardiac surgery was implanted with an impella 5.5 device for mechanical circulatory support.The complainant reported that complications were encountered during implant of of an impella 5.5 pump.When advanced into place, the.018" guidewire was seen under echocardiogram (echo) to be located under the papillary muscle.The decision was made to remove the pump and guidewire for re-insertion; however, the guidewire was found to be kinked.The wire was swapped and the pump was implanted without further issue.After implant, some oozing was noted at the axillary access site, but intervention was not required.There was no patient harm and support with the implanted pump continued as planned.
|