BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4482 |
Device Problems
Fluid/Blood Leak (1250); Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the twist clamp of a minicap transfer set ¿were not closing¿ which resulted in a leak.This was observed prior to use of the device for peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The device was received for evaluation.A visual inspection with the naked eye identified a separation between the dark blue female connector and the light blue main body of the twist clamp.Functional testing including leak, clear passage and clamp function testing were performed with no issues noted.The reported condition was verified.The cause of the condition was determined to be manufacturing related due to inadequate solvent application to the main body of the twist clamp.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
|