• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT TENDRIL STS; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT TENDRIL STS; NO MATCH Back to Search Results
Model Number 2088TC/58
Device Problems Intermittent Capture (1080); Material Twisted/Bent (2981)
Patient Problem Syncope/Fainting (4411)
Event Date 09/18/2023
Event Type  Injury  
Event Description
It was reported, the patient presented in the emergency room due to syncope.Upon interrogation, intermittent capture was observed on the right ventricular (rv) lead.Diagnostic imaging was performed and a compression of the rv lead was observed.The rv lead was explanted and replaced to resolve the event.The patient was stable.
 
Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17862949
MDR Text Key324828750
Report Number2017865-2023-47627
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502894
UDI-Public05414734502894
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2088TC/58
Device Lot NumberP000145627
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-