Model Number 2090
Device Problems Contamination (1120); Computer Operating System Problem (2898)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2023
Event Type  malfunction  
Event Description
It was reported that the programmer was slow to start, restarted during boot up and was unable to interrogate implantable devices.Changing out the radiofrequency (rf) head failed to resolve the issue.An alternative programmer was used to perform the interrogation.The programmer was returned for evaluation and subsequently tested out of specification during manufacturer's analysis.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis: it was determined on analysis that there was internal corrosion on multiple parts of the micro processor unit (mpu) board and it was worn.Analysis was able to partially confirm the customers comment that the programmer was slow to start and restarted during boot up; it was noted that the power supply was defective.It was further noted that the keyboard hinge was broken, the lower bullets were worn, the bezel was broken round the edges (considered acceptable), key on the keypad was loose and the upper/lower case had cosmetic damage.All found defective parts were replaced and all other identified issues were resolved.Reconfigured hard drive, reloaded and updated software and the programmer then passed all final functional and systems tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key17867227
MDR Text Key324877326
Report Number2182208-2023-02833
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 10/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
229047 ANALYZER, 2067 RADIOFREQUENCY HEAD
-
-