It was reported a patient presented for an implant procedure on (b)(6) 2023.During the operation, the right ventricular (rv) lead presented with r-wave amplitude variation due to partial lead dislodgement.The lead was not used and replaced within the same operation, it was also noted the lead was bumpy and the insulation was twisted.Post-procedure the patient was stable.
|
The reported events were lead dislodgement, r-wave amplitude variation and ¿lead was bumpy, and insulation twisted¿.As received, a complete lead was returned in one piece with the helix found retracted and clogged with dried blood/tissue.After cleaning, the helix can be extended and retracted by applying torque directly at the connector pin.The measured full helix extension length was within specification.The reported event of r-wave amplitude variation was not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.The reported event of ¿lead was bumpy, and insulation twisted¿ was not confirmed.Visual and x-ray inspections of the lead did not find any anomalies.
|