Lot Number VMFD290523
Device Problems Device-Device Incompatibility (2919); Difficult to Advance (2920); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2023
Event Type  malfunction  
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event Description
It was reported during a watchman left atrial appendage closure (laac) procedure, a versacross connect kit was selected for use.The physician gained access with versacross and watchman sheath.However, when tried to move the mechanical guidewire back and forth, it could not.A resistance was felt when inserting the wire into the dilator into the superior vena cava (svc).Hence the wire was pulled out with the dilator tip.As a resolution, versacross radio frequency wire was used instead.The procedure was completed successfully and no patient complications occurred.The device is not expected to be returned for analysis as it was contaminated.It was further advised, nothing was off about patient anatomy and nothing looked strange upon opening the connect system.Every time it get used it gets stuck in the dilator and have physician to remove the dilator and pull the j wire out from the tip of the dilator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACROSS CONNECT LAAC ACCESS SOLUTION
Type of Device
DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
Manufacturer (Section D)
BAYLIS MEDICAL COMPANY INC.
5959 trans-canada highway
montreal, qc H4T 1 A1
CA  H4T 1A1
Manufacturer (Section G)
BAYLIS MEDICAL COMPANY INC.
5825 explorer drive
mississauga, on L4W 5 P6
CA   L4W 5P6
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key17882154
MDR Text Key325076877
Report Number2124215-2023-54635
Device Sequence Number1
Product Code DRE
UDI-Device Identifier00685447012597
UDI-Public00685447012597
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K220414
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot NumberVMFD290523
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2023
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
-
-