• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; NO MATCH Back to Search Results
Model Number 2088TC/52
Device Problem Material Twisted/Bent (2981)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that the lead was twisted or bent.No intervention or patient condition was reported.
 
Manufacturer Narrative
The reported event of ¿lead appears slightly kinked¿ was confirmed.A complete lead was returned in one piece with the helix found retracted and clogged with dried blood/tissue.Visual inspection found the outer coil bent at the distal region.The cause of the reported event was isolated to outer/inner coils bent consistent with procedural damage.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the lead did not find any anomalies except for procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17882708
MDR Text Key325074144
Report Number2017865-2023-47440
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502887
UDI-Public05414734502887
Combination Product (y/n)Y
Reporter Country CodeCA
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2088TC/52
Device Lot NumberP000170952
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-