The mayfield skull clamp ((b)(4)) was returned for evaluation: device history record (dhr) - the dhr was reviewed, and no anomalies related to the reported failure was observed.Failure analysis - the investigation of the returned device showed that the reported complaint was confirmed from the evaluation.The skull clamp had rotational movement in its swivel lock assembly and a residue buildup was present.For the adjustment, some worn-off internal parts required replacement, and general maintenance and cleaning were required.To exclude the ratchet from opening it was inspected specifically; the ratchet extension showed a displaced helicoil insert and there was a sudden movement when trying to apply pressure onto the torque screw.The ratchet required complete replacement along with the plunger stud.To resolve all issues, the internal parts to adjust the unit according to the manufacturer were replaced, including the ratchet extension and the plunger stud, and the skull clamp's swivel lock assembly was cleaned.After completing the reassembly, the skull clamp was subjected to a successful function test.Root cause - the complaint is confirmed via inspection of the unit.Probable root cause of the observed device deficiencies is routine use and rough handling.Additionally, improper, or suboptimal placement of the skull clamp can contribute to patient movement.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|