The mayfield skull clamp ((b)(6)) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the reported complaint was confirmed from the evaluation.The skull clamp has a loose swivel lock, and a residue buildup was present.Some worn off internal parts needed to be replaced along with the swivel base, which could have a worn off inner thread, and the torque screw was mixed up and marked with the instance no.(b)(4).As a result, the unit was scrapped.Root cause - complaint is confirmed via inspection of the unit.Probable root cause is routine use and wear.Additionally, improper or suboptimal placement of the skull clamp can contribute to movement of the patient.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|