Catalog Number 2C2063K |
Device Problems
Filling Problem (1233); No Flow (2991)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a large volume infusor would not flow.This was observed after filling with 6g cefazolin.When the blue winged cap was removed, no liquid flowed out.A syringe was used to dislodge any particle blocking the flow.Flow was subsequently established; however, the flow was slower than normal.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H4: the lot was manufactured on november 14, 2022 ¿ november 15, 2022.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional flow rate test was performed, and the results were found to be within the product specification range.Based on the sample evaluation result, the infusor unit was determined to be conforming product.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|