OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problems Intermittent Loss of Power (4016); Unexpected Shutdown (4019)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Other text: b3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the device's power switch was faulty.When it was switched on it immediately turns off.If there is slight pressure on the switch it will stay on.Patient involvement is unknown.
 
Manufacturer Narrative
Email is: (b)(6).Device evaluation: no product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.The exact cause could not be determined; however, based on the reported problem, the most probable cause would be an outdated pcb board with a broken power switch leads connection to the board.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.If the product is returned, the manufacturer will reopen this complaint for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17886653
MDR Text Key325850971
Report Number3012307300-2023-09563
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public10695085002796
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberHL-390
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/31/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-