For the investigation the logfile and the replaced therapy control unit were analysed.On the reported date of event the descriped reboot could be reconstructed.Before and after the reboot ventilation was stable and unremarkable.The affected pcb was tested in the laboratory and the problem could be confirmed.In case of a reset of the processors of the therapy control unit therapy is discontinued for a maximum of 15 seconds, before therapy will be resumed with the latest valid settings.If for any reasons therapy was not restarted within 15 seconds an acoustical alarm will be given.In the present case the system behaved according to specification for this failure case.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.The field failure rate is subject to permanent monitoring by the responsible product quality board and is rated as acceptable.
|