The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was confirmed.The image was found intermittently when the unit was in the ¿up¿ position.Additionally, the biopsy channel had a leak present, and the insertion tube as well as the adhesive were both compromised.If additional information becomes available following the device evaluation, a supplemental report will be filed.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, a definitive root cause of the reported image loss upon angulation issue could not be determined, however, the issue was likely due to physical stress applied to the insertion site during user handling/mishandling, resulting in damage to the image sensor the image sensor unit.The event can be detected/prevented by following the instructions for use (ifu) which states: important information ¿ please read before use precautions: caution turn the video system center on only when the endoscope connector is connected to the light source.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.Precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system inspection of the endoscopic image confirm that the wli and nbi endoscopic images are normal.5.1 troubleshooting if any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2,¿troubleshooting guide¿.If the problem still cannot be resolved, send the endoscope to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Also, should any irregularity be observed while using the endoscope, stop using it immediately and withdraw the endoscope from the patient as described section 5.3, ¿withdrawal of the endoscope with an irregularity¿.Olympus will continue to monitor field performance for this device.
|