MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
High impedance (1291); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/27/2023 |
Event Type
malfunction
|
Event Description
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was being implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that there were high impedances of 40,000 ohms seen during the procedure on electrodes 5-9.There were no contributing factors.The surgeon has cleaned them, detached from the ins.The neurostimulator was clean.The lead was also clean, blood was removed.Troubleshooting was performed. impedances tests multiple times, session was shut down, and the tablet was shut down too.The lead was replaced and the issue was resolved.No surgical intervention occurred, nor was planned.The patient was alive with no injury.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 977c165 lot# serial# (b)(6) implanted: (b)(6) 2023 explanted: (b)(6) 2023 product type lead section d information references the main component of the system.Other relevant device(s) are: product id: 977c165, serial/lot #: (b)(6), ubd: 26-jul-2027, udi#: (b)(4).H6 codes belong to the lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3.Device analysis for lead va2u1fq017 revealed no anomaly.H6 codes belong to the lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|