SMITH & NEPHEW, INC. TWINFIX ULTRA PK 4.5MM W/2 UB -WHT &BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
Catalog Number 72202595 |
Device Problems
Break (1069); Fracture (1260); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/22/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that, during a medial collateral ligament surgery, the screw of the twinfix ultra broke while it was being screwed in.All the pieces were removed from the patient with surgical tools.Surgery was completed after a non-significant delay, using a back-up device in the originally drilled bone hole, no void was left.No further complications were reported.Patient's status is good.
|
|
Manufacturer Narrative
|
H10: internal complaint reference: (b)(4).
|
|
Manufacturer Narrative
|
H10: h3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The insertion device was returned with the fractured anchor and suture strings.The anchor is fractured from the proximal end of the suture window toward the base.The prongs at the distal end of the insertion shaft are deformed.There is biological debris on the returned items.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the raw material strength requirements and storage requirements are specified and each lot must be accompanied by a material certificate of analysis.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|