It was reported that the procedure was to treat a heavily calcified, heavily tortuous left main artery.The 3.50x33mm xience skypoint stent delivery system (sds) was implanted; however, noted to have an issue of deliverability, overall performance and positioning accuracy.There was no adverse patient effects and no clinically significant delay.No additional information was provided.
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|