• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number SONATA
Device Problem Impedance Problem (2950)
Patient Problems Ossification (1428); Failure of Implant (1924)
Event Date 08/29/2023
Event Type  malfunction  
Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
The recipient's hearing etiology is otosclerosis and osteogenesis.However, it was indicated that the hearing performance with the device was not affected.Further proceedings are unknown.
 
Manufacturer Narrative
Additional information: according to the information received from the field in situ measurements show six channels with fluctuating high impedance likely caused by physiological changes in the cochlea due to the reported otosclerosis and osteogenesis.No information on possible further steps has been received.
 
Event Description
The recipient's hearing etiology is otosclerosis and osteogenesis.However, it was indicated that the hearing performance with the device was not affected.Further proceedings are unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key17904053
MDR Text Key325271484
Report Number9710014-2023-00906
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/01/2014
Device Model NumberSONATA
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/02/2023
Date Device Manufactured10/01/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexFemale
-
-