VYAIRE MEDICAL INFANT FLOW LP NCPAP GENERATOR SET; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
|
Back to Search Results |
|
Model Number PACKAGE ASSEMBLY,INFANT FLOW,LP |
Device Problems
Partial Blockage (1065); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/01/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
H3: 81 other - the sample has not been returned for evaluation.Therefore, no root cause determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
It was reported to vyaire medical that the 7772020lp package assembly, infant flow, lp has excessive rain discharge that forms within the exhaust limb and pools at the patient interface, significantly irritating neonatal patients and posing a serious risk of water entering the patient's airway.Furthermore, there is no information regarding patient harm.
|
|
Manufacturer Narrative
|
Results of investigation: vyaire medical was not able to verify the reported issue.The suspect device was not returned for evaluation.Investigation findings are derived from r&d engineering investigation report (45-r5157 v.01) of rainout performance between vyaire medical infant flow-low pressure (iflp) generator and the inspire generator.The r&d team conducted several test cases to assess, verify, and understand the different scenarios of rainout as well as assess overall performance between the two generators.The r&d team repeated the testing performed by cumh.No condensate was observed in the inspiratory line or generator in both the original cumh testing or the vyaire testing.The amount of condensate in the exhaust tubing was found to be influenced by incorrect set-up.Incorrect placement of temperature probe inside the incubator was found to increase the condensate found in the iflp exhalation tube.Also, horizontal positioning of the exhalation tubing was found to increase the condensation collection in the exhalation tubing.In all testing, there was some condensation in the exhalation tubing for both the iflp generator and the inspire product.Next, the r&d team performed further testing to determine if rainout could be further reduced without causing detrimental performance.Testing with the exhalation tube slits variously closed showed some variation in the amount of condensation; however, no configuration performed consistently better than the un-modified device.In measuring the temperature of the gas in the exhalation tubing, iflp was shown to deliver a more consistent and closer humidifier temperature setting (37°c and 44mg/l) to the patient than the inspire products.The r&d team concluded their investigation none of their findings suggest that the current device (iflp) produces more condensation in the exhaust limb than either of the previous vyaire (cah) products or the initial release of the iflp; furthermore, condensation is very heavily influenced by setup, specifically with the positioning/draping of the exhalation tubing.There was no device problem detected.The amount of condensate observed in the exhaust tubing is influenced heavily by the set-up of the iflp circuit.Following the guidance of f&p and vyaire medical may reduce the amount of condensate in the exhaust line.While there was condensate in the exhaust portion of the iflp circuit, there was no condensate in the inspiratory line or generator observed in the engineering investigation.Based on the information gathered, iflp design performs as intended with an acceptable level of condensation in the exhaust line.
|
|
Search Alerts/Recalls
|
|
|