Catalog Number 261221
Device Problem Fail-Safe Did Not Operate (4046)
Patient Problem Laceration(s) (1946)
Event Date 09/25/2023
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A nurse reported a perforator (id 261221) did not stop during the second cranial hole, causing dural damage.The incident led to surgical delay because hemostatic agent had to be used to stop the bleeding.The perforator was tested before the surgery.No additional information has been provided after several attempts.
 
Event Description
N/a.
 
Manufacturer Narrative
Perforator (id 261221) was returned for evaluation.Failure analysis - the perforator unit was inspected using the unaided eye: the unit was heavily soiled with organic material between the inner and outer drill and had a worn label.Instructions for use (ifu) testing procedure was performed after freeing the inner and outer drill by drilling test holes.The organic matter had to be freed before testing.Once freed, the unit passed the spring test as designed.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release.The unit was found to perform as intended and fulfilled the acceptance criteria.It is suspected that the organic material caused some friction and difficulty at first, but as more organic material was freed, drilling was found to occur as expected.Root cause analysis - the root cause is undetermined and was unable to be confirmed in the complaint evaluation.Product was received for analysis and the investigation could not confirm the complaint.Potential causes of failure include, user misuse.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17907252
MDR Text Key325340800
Report Number3014334038-2023-00169
Device Sequence Number1
Product Code HBF
UDI-Device Identifier10381780513599
UDI-Public10381780513599
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-