Model Number DDBB2D1
Device Problems False Alarm (1013); Electromagnetic Interference (1194); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 10/05/2023
Event Type  malfunction  
Event Description
It was reported that a lead integrity alert (lia) triggered due to high rate non-sustained episodes and elevated sensing integrity counter (sic).It was noted that the patient was welding at the time.The implantable cardioverter defibrillator (icd) was suspected of sensing electromagnetic interference (emi) from the welding.It was stated that the patient has a history of welding that triggers lia alerts, of which the first occurred in (b)(6) 2017 where the patient received an inappropriate shock due to the oversensing during welding.Additionally, a review of the data/episodes shows fine chatter during periods of oversensing consistent with patient emi exposure.It was also reported that at the time of an in office evaluation with the patient, oversensing was observed on the right ventricular (rv) lead during impedance testing; there had been no non-physiological oversensing up until then. the icd and rv lead remain in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Continuation of d10: 6943 lead implant date: (b)(6) 2004 and 5076 lead implant date: (b)(6) 2004.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that reprogramming was done.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERA¿ XT DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key17908172
MDR Text Key326251943
Report Number9614453-2023-03622
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P980016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/28/2019
Device Model NumberDDBB2D1
Device Catalogue NumberDDBB2D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/15/2023
Date Device Manufactured10/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10...
Patient Outcome(s) Required Intervention; Life Threatening;
-
-