• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G447
Device Problems Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Inaccurate Synchronization (1609)
Patient Problems Arrhythmia (1721); Loss of consciousness (2418); Electric Shock (2554)
Event Date 09/27/2023
Event Type  Injury  
Event Description
It was reported that electrogram (egm) review of this cardiac resynchronization therapy defibrillator (crt-d) system was requested to determine whether tachy therapy was delivered and therapy was exhausted in two stored episodes due to true ventricular tachycardia (vt) or atrial fibrillation (af) with rapid ventricular response (rvr).Additionally, it was unclear whether the first shock initiated vt, and it was suspected that anti-tachycardia pacing (atp) had accelerated the patient's rhythm.Technical services (ts) reviewed troubleshooting options and programming considerations.The field representative was unsure whether the patient had lost consciousness.It was noted that the patient was in chronic af, the patient currently had covid-19, and the patient's heartlogic index score was 46.This crt-d system remains in service.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17908704
MDR Text Key325335571
Report Number2124215-2023-56115
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589539
UDI-Public00802526589539
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2022
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number235462
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/10/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age74 YR
Patient SexMale
-
-