• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PERSEUS A500; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PERSEUS A500; ANESTHESIA UNITS Back to Search Results
Catalog Number MK06000
Device Problems Circuit Failure (1089); Gas Output Problem (1266); Failure to Deliver (2338)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2023
Event Type  malfunction  
Event Description
It was reported there was a vent fail with alarms.No patient injury reported.
 
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.H3 other text : on-going.
 
Manufacturer Narrative
According to the logfile records the internal 12 v/24 v supply got temporariliy interrupted for an unknown reason.The device alarmed for turbovent2 failure, ventilator failure and system failure.All relevant internal voltages are generated from the 24 v supplied by the power supply unit on the pba main a500 and provided to the connected pcbs.Thus, root cause was a faulty pba main a500.The exact root cause couldn¿t be determined, most likely one of the voltage regulators was faulty.The device issued appropriate alarms.
 
Event Description
It was reported there was a vent fail with alarms.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSEUS A500
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key17910880
MDR Text Key325359389
Report Number9611500-2023-00357
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K133886
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMK06000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/31/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-