• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT CRM S.R.L. KORA; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT CRM S.R.L. KORA; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number KORA 100 DR
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2023
Event Type  malfunction  
Manufacturer Narrative
Conclusion: the programmer couldn¿t identify the device because of interferences when the device was interrogated with the cpr4 header.Recommendations: please check the presence of other machines in the room and when possible, switch them off.The kora pacemaker is not involved in the difficulty to interrogate him with the cpr4 header.
 
Event Description
Reportedly, interrogating a kora pacemaker, the identification of the device wasn't possible (4 blue lights).The tablet smarttouch was restarted (long press on the off/on button).The same situation happened again.The pacemaker was interrogated with a second tablet (+ cpr3 / xb tablet).This happened again with other implantable devices (difficulties to interrogate with the cpr4 head).The cpr4 head is connected to the usb port of the tablet, the tablet is on its basis, connected to the power.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KORA
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT  13040
Manufacturer (Section G)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT   13040
Manufacturer Contact
elodie vincent
via crescentino s.n
saluggia (vc) 13040
IT   13040
146013665
MDR Report Key17913320
MDR Text Key325404324
Report Number1000165971-2023-00831
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberKORA 100 DR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient EthnicityNon Hispanic
-
-