Model Number 2110
Device Problems Defective Alarm (1014); Air/Gas in Device (4062)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported the device didn't administer medication for five hours.The disposable was not fully spiked so no medication could enter the tubing.However, the pump never alarmed for air or for a proximal occlusion.No adverse patient effects were reported by the customer.
 
Manufacturer Narrative
Other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.B3: unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD SOLIS HPCA
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17916459
MDR Text Key326255594
Report Number3012307300-2023-09619
Device Sequence Number1
Product Code MEA
UDI-Device Identifier15019517154993
UDI-Public15019517154993
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2110
Device Catalogue Number21-2112-0402-51
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/07/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-