Model Number 6300
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/11/2023
Event Type  malfunction  
Manufacturer Narrative
Other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the pump was left to run and it was later noticed that the pump bag, battery pack and display screw were wet.No patient injury.
 
Manufacturer Narrative
Other, other text: no product was returned.We are unable to confirm the reported complaint.If the product is returned, the manufacturer will reopen this complaint for further investigation.The most probable cause for a pump bag being wet is a leak in the cassette or tubing.The most probable cause for battery pack and lcd screen being wet is fluid ingression; however, this cannot be confirmed as no product was returned for investigation.The service history review identified no indication that the complaint was related to a service of the device within the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD LEGACY PCA AMBULATORY INFUSION SYSTEM
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17916526
MDR Text Key326255446
Report Number3012307300-2023-09622
Device Sequence Number1
Product Code MEA
UDI-Device Identifier10610586019449
UDI-Public10610586019449
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K982839
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6300
Device Catalogue Number21-6300-51
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/19/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-