MEDOS INTERNATIONAL SARL SHAVER HANDPIECE 2.0 WITH BUTTONS; ARTHROSCOPE Back to Search Results
Catalog Number 283812
Device Problems Overheating of Device (1437); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported by the affiliate in sweden that during an unspecified procedure on an unknown date, the shaver handpiece 2.0 device sounded bad and became hot.Another like device was used to complete the procedure.There were no delays reported.There were no adverse patient consequences reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation.Therefore, the reported condition cannot be confirmed and/or duplicated.H4: the device manufacture date is currently unavailable.Udi: (b)(4).
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: investigation summary : the complaint device was received at the r&d.The unit involved in this complaint has been received and per s&r technicians, these units fall under capa and escalation, therefore no service/wo will be performed.These units presented a failure in the motor; however, as the motor cannot be disassembled to perform a repair, the devices must be discarded.The front seal of the motor is allowing saline to enter the motor during use.This causes the handpiece to stop functioning.This fault affects the reliability of the handpiece but does not adversely impact patient safety.Manufacturing record evaluation (mre) is not needed since the device falls into capa that was opened due to the motor cannot be disassembled to perform a repair.The products (283712 & 283812) produced with the original design are not serviceable.The root cause is attributed to manufacturing involving a design change.There is a capa associated with this product code, most of the complaints received to-date are related to failed motors with evidence of water intrusion and manifested as malfunctioning or seized motors, therefore; the devices are under service discontinuation.This product issue is already being addressed by depuy quality system.Further investigation and assessment are covered under the capa in our quality system.Depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHAVER HANDPIECE 2.0 WITH BUTTONS
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key17919860
MDR Text Key325489863
Report Number1221934-2023-03701
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705030743
UDI-Public10886705030743
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K041824
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number283812
Device Lot NumberM28AB2019
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/21/2023
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-