This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, since the device was not returned for evaluation, no abnormalities were confirmed.Therefore, the root cause of the event could not be determined.However, the event was likely caused by the following: 1a person assisting with the procedure pulled the slider too much, and the polyp was mechanically resected.The polyp resection was performed mechanically, which caused the bleeding.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿the operator of this instrument must be physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical endoscopic technique.This manual, therefore, does not explain or discuss clinical endoscopic procedures.¿ ¿operation of this instrument is based on the assumption that open surgery is possible as an emergency measure if the loop cannot be detached from the instrument or if any other unexpected circumstance takes place.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ in this manual.¿ ¿use this instrument in an environment equipped to accommodate open surgery and have the hospitalization plan prepared in case any problem occurs that may not be resolved endoscopically.¿ ¿never use excessive force to operate the instrument.This could damage the instrument.¿ ¿do not insert the instrument into the endoscope unless you have a clear endoscopic field of view.If you cannot see the distal end of the insertion portion in the endoscopic field of view, do not use it.This could cause patient injuries, such as punctures, hemorrhages, or mucous membrane damage.It may also damage the endoscope and/or instrument.¿ ¿do not withdraw the instrument or change the angulation of the endoscope before the loop is detached from the instrument.Doing so may tear tissue inside the body cavity, resulting in patient injury, such as punctures, hemorrhages, or mucous membrane damage.¿ ¿do not apply unnecessary force to the loop when ligating tissue during the procedure.Excessive force applied to the loop could cut the surrounding body cavity tissue, resulting in patient injury, such as hemorrhages or mucous membrane damage.It may also damage the endoscope and/or instrument.¿ this supplemental report includes a correction to b1 and g2.¿product problem¿ was inadvertently selected on the initial medwatch.This has been corrected to "adverse event" only, as there was no reportable device malfunction.Also, information has been added to g2 that was inadvertently not included on the initial medwatch.Olympus will continue to monitor field performance for this device.
|