• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BRAEMAR MANUFACTURING, LLC C6 MCOT PPM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BRAEMAR MANUFACTURING, LLC C6 MCOT PPM Back to Search Results
Model Number 02-01894
Device Problems Melted (1385); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2023
Event Type  malfunction  
Event Description
It was reported that the patient starting charging the monitor however the monitor charger melted.The monitor was damaged and the charging cord was very difficult to remove from the device.The patient stopped using the device and a replacement was sent.The patient stated that there was no harm or property damage.
 
Manufacturer Narrative
It was reported that the monitor charger melted and caused damage to the monitor.The device was returned for investigation.Engineering evaluation was able to confirm device melt.No internal device was observed during evaluation.It is most probable that the device melt was caused by an external heat source.Am&d philips is doing further investigation fro mcot usb-a/usb-c cables overheating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C6 MCOT PPM
Type of Device
C6 MCOT PPM
Manufacturer (Section D)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer (Section G)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer Contact
beverly okoh
1285 corporate center drive
suite 150
eagan, MN 55121
MDR Report Key17920182
MDR Text Key325493319
Report Number2133409-2023-00062
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number02-01894
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received01/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age30 YR
Patient SexFemale
-
-