MEDOS INTERNATIONAL SARL TRUESPAN 12 DEGREE PEEK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 228151
Device Problems Degraded (1153); Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.H4: the device date of manufacture is unknown at this time.E3: reporter is a j&j sales representative.Udi: (b)(4).
 
Event Description
It was reported by the sales rep that during an unspecified procedure on (b)(6) 2023 the truespan 12 degree peek device became stuck and it was not possible to recover the point.It was reported that the red trigger was loose and not fixed as usual.Another like device was used to complete the procedure.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: a photo was returned to depuy synthes mitek for evaluation.The depuy synthes mitek team conducted a visual inspection of the returned photo.Upon visual inspection of the photo, it was noted that the device trigger was fully retracted to the handle which means that the internal trigger mechanism that retract the trigger to its original position is broken.A manufacturing record evaluation was performed for the finished device, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.Based on the visual findings, the reported complaint was confirmed.A possible root cause can be attributed to a portion of tissue that was entered inside the applier needle causing a mechanical obstruction of the plate when it was going to be deployed; this obstruction and the force applied to the trigger are contributive factors of the broken trigger, however this cannot be conclusively determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of the device received.The complaint device was received and evaluated; visual inspection reveals that the red trigger is broken.The plates and suture were not returned.The device was opened to verify the broken part, it was found that the upper part of the red trigger was broken.The applier needle was found in a normal condition.A manufacturing record evaluation was performed for the finished device lot number: 9l92792, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection, this complaint can be confirmed.The possible root cause for the issue experienced by the customer can be attributed to a portion of tissue that was entered inside the applier needle causing a mechanical obstruction of the first plate when it was going to be deployed; this obstruction and the force applied to the trigger are contributive factors of the broken trigger, however this cannot be conclusively determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance device history review
=
> pn:(b)(6) lot number:9l92792 there was no non conformance regarding this lot manufacturing date: 12 oct 2022 expiry date: 30 sep 2025 quantity: 100ea release date: 28 oct 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN 12 DEGREE PEEK
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raymham, MA 02767
3035526892
MDR Report Key17921871
MDR Text Key325511242
Report Number1221934-2023-03708
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026012
UDI-Public10886705026012
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number228151
Device Lot Number9L92792
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received12/05/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexMale
-
-