ROCHE DIAGNOSTICS ACCU-CHEK INFORM II TEST STRIPS; BLOOD GLUCOSE MONITORING TEST STRIPS Back to Search Results
Catalog Number 05942861001
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/25/2023
Event Type  malfunction  
Event Description
There was an allegation of questionable glucose results from an accu-chek inform ii meter.Patient 1: at 11:57 am, the initial result was hi (>600 mg/dl).At 12:04 pm, the retest result was 196 mg/dl.Patient 2: at 7:19 am on (b)(6)2023, the initial result was 480 mg/dl.At 7:45 am, the retest result was 48 mg/dl.
 
Manufacturer Narrative
The accu-chek inform ii meter serial number was (b)(6).Quality control results on the meter were acceptable.On a regular basis, accu-chek inform strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.The customer material was requested for further investigation.
 
Manufacturer Narrative
None of the customer's materials were received for investigation.Therefore, further investigation was not possible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-CHEK INFORM II TEST STRIPS
Type of Device
BLOOD GLUCOSE MONITORING TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17925356
MDR Text Key326253076
Report Number1823260-2023-03280
Device Sequence Number1
Product Code LFR
UDI-Device Identifier00365702428102
UDI-Public00365702428102
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K121679
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05942861001
Device Lot Number670221
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-