• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4087K
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that a large volume folfusor had no flow; the device did not work for seven (7) days.This was discovered during patient use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the device was received for evaluation containing approximately 202ml of fluid in the bladder.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional flow rate test was performed and the flow rate was found to be within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17926809
MDR Text Key325589426
Report Number1416980-2023-05182
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412460987
UDI-Public(01)00085412460987
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4087K
Device Lot Number21N034
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received10/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-