Catalog Number 702-04-58F
Device Problem Corroded (1131)
Patient Problem Pain (1994)
Event Date 09/19/2023
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.H3 other text : not returned to the manufacturer.
 
Event Description
It was reported that the patient's right hip was revised.2 months pos-op, patient complaint of increasing pain.Approximately 1 month prior to revision, increased ion levels were detected.Intra-operatively, corrosion was noted on the mdm metal liner between the metal liner and the shell.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENTII TRITANIUM CLUSTER58F
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17928632
MDR Text Key325561777
Report Number0002249697-2023-01189
Device Sequence Number1
Product Code MEH
UDI-Device Identifier07613327344028
UDI-Public07613327344028
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K171768
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 10/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number702-04-58F
Device Lot Number96596301A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age66 YR
Patient SexMale
-
-