• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CARBOFLO; EPTFE VASCULAR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CARBOFLO; EPTFE VASCULAR GRAFT Back to Search Results
Catalog Number 40S08C
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/19/2023
Event Type  malfunction  
Event Description
It was reported that during a surgical graft placement procedure, the graft was allegedly found to be split.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiration date: 06/2027).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device pending return.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: received one impra carboflo graft segment for evaluation.The sample appeared to have blood residue throughout.A longitudinal split was noted on the one end of the graft.Upon visual evaluation, the graft has been cut on both ends.Crimp marks were noted on end one.A large tear was noted approximately 16mm from end two.Therefore, the investigation is determined to be confirmed for the reported material split, cut or torn issue as split was noted on the end of the graft.A definitive root cause for the reported material split, cut or torn issue could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: b5, d4 (expiration date: 06/2027), g3.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a surgical graft placement procedure, the graft was allegedly found to be splitting when sutures were placed.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBOFLO
Type of Device
EPTFE VASCULAR GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17929050
MDR Text Key326270174
Report Number2020394-2023-00859
Device Sequence Number1
Product Code DSY
UDI-Device Identifier00801741020810
UDI-Public(01)00801741020810
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K004011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number40S08C
Device Lot NumberVTGT0348
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-