C.R. BARD, INC. (COVINGTON) -1018233 DAVOL® PEDIATRIC CATH KIT Back to Search Results
Catalog Number 0035640
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2023
Event Type  malfunction  
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿components damage ¿ received from supplier".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "infant and pediatric cath kits: information for use: caution: read all instructions prior to use.Indications: bladder catheterization is performed when a sterile urine sample is needed and a clean catch specimen cannot be obtained.Warnings: the catheter should never be forced if resistance is felt.Trauma to the bladder (false passage) can be prevented by adequately lubricating the catheter and stretching the penis to straighten the urethra.Sterile: sterile unless package is opened or damaged.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Instructions: 1.Open plastic wallet, which may be used as a sterile field.The following four steps apply only to male catheterization: 2.Place the infant supine, with the thighs abducted (frog-leg position).3.Cleanse the penis with povidone-iodine swabs, starting with the meatus and moving in a proximal direction.4.Put on sterile gloves.Remove protective sheath from catheter and pull catheter out of centrifuge tube to desired length.Lay tube assembly on the sterile field.5.Place the tip of the catheter in sterile lubricant.Hold the penis perpendicular to the body to straighten the penile urethra and help prevent false passage.Advance the catheter carefully until urine appears.A slight resistance may be felt as the catheter passes the external sphincter.Note: if urine does not flow freely into the tube, the cap may need to be loosened slightly.6.After urine is collected, pull catheter out of the cap of the centrifuge tube.Tighten cap and depress spout.7.Fill out label and place on centrifuge tube.Send to lab in normal manner.Instructions: 1.Open plastic wallet, which may be used as a sterile field.The following four steps apply only to female catheterization: 2.Place the infant supine, with the thighs abducted (frog-leg position).3.Separate the labia and cleanse the area around the meatus with the povidone-iodine swabs.Use anterior-to-posterior strokes to prevent fecal contamination.4.Put on sterile gloves.Remove protective sheath from catheter and pull catheter out of centrifuge tube to desired length.Lay tube assembly on the sterile field.5.Place the tip of the catheter in sterile lubricant.Spread the labia with two fingers and carefully advance the catheter until urine appears.Note: if urine does not flow freely into the tube, the cap may need to be loosened slightly.6.After urine is collected, pull catheter out of the cap of the centrifuge tube.Tighten cap and depress spout.7.Fill out label and place on centrifuge tube.Send to lab in normal manner." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the physician ordered a catheter to be catheterized on the 10 months old patient.They prepared the area, inserted the catheter, the catheter broke away from the plastic cup that catches the urine.No harm to the patient.Several other catheter kits with the same lot number were opened and were noted to have the same concern.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVOL® PEDIATRIC CATH KIT
Type of Device
DAVOL® PEDIATRIC CATH KIT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17930327
MDR Text Key325604389
Report Number1018233-2023-07418
Device Sequence Number1
Product Code FFH
UDI-Device Identifier00801741045851
UDI-Public(01)00801741045851
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0035640
Device Lot NumberNGGV0947
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-