US SURGICAL PUERTO RICO ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIA60AMT
Device Problems Entrapment of Device (1212); Retraction Problem (1536); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Tissue Breakdown (2681)
Event Date 09/19/2023
Event Type  malfunction  
Manufacturer Narrative
D10 concomitant products: scd7a39, sonicision 7 dissector scd7a39 39 cm, (lot #30190030x); egiaustnd, egiaustnd endogia ultra univ std stap, (lot #p3d0463).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, intra-operatively of a laparoscopic restorative proctectomy, when attempting to resect the area of disease in the lower rectum, the device fired the staples but the surgeon was then unable to retract the blade or open the jaws after firing.The stapler was locked closed on the large bowel and would not open.The surgeon subsequently had to change the plan and take a perineal approach in order to release the locked stapler.The surgical time was extended for 30 minutes or more.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the reload was fully fired, the jaws of the reload were closed, and the instrument was received attached to the listed instrument.The reload knife laminates were found to be damaged.Functionally, the reload knife was forcefully retracted.The reload was removed from the instrument.It was reported that the instrument locked on tissue and was difficult to retract knife blade.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17931773
MDR Text Key325624193
Report Number2647580-2023-04426
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884523003215
UDI-Public10884523003215
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEGIA60AMT
Device Catalogue NumberEGIA60AMT
Device Lot NumberP3D0137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10
Patient Age71 YR
Patient SexMale
Patient Weight116 KG
-
-