• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS CREP2; ENZYMATIC METHOD, CREATININE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS CREP2; ENZYMATIC METHOD, CREATININE Back to Search Results
Catalog Number 05168589190
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2023
Event Type  malfunction  
Event Description
There was an allegation of a questionable creatinine (crep gen.2) result from the cobas 8000 c702 module.The initial result was 9.7 mg/dl and was reported outside of the laboratory.The patient had a new sample drawn at another facility and the result did not match.No specific data was provided.The original sample was repeated and the result was 0.6 mg/dl.This result was believed correct.
 
Manufacturer Narrative
The analyzer serial number was (b)(6).The investigation is ongoing.
 
Manufacturer Narrative
The analyzer alarm trace contained multiple sample short and abnormal aspiration alarms on the date of the event near the time the sample was processed.The field service engineer found a film in the water tank.He performed a full decontamination and ran precision testing.The customer ran qc and all results met system specifications.Operational and/or mechanical checks passed.The investigation determined the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CREP2
Type of Device
ENZYMATIC METHOD, CREATININE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17932058
MDR Text Key326048115
Report Number1823260-2023-03294
Device Sequence Number1
Product Code JFY
UDI-Device Identifier04015630924998
UDI-Public04015630924998
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K024098
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number05168589190
Device Lot Number71018001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/19/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
-
-