Model Number BF-P190 |
Device Problems
Leak/Splash (1354); Mechanical Problem (1384); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/19/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus that the evis exera iii bronchovideoscope had angulation malfunction, the light guide lens had defects, and a b30 scope communication error.There were no reports of patient or user harm associated with this event.
|
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation and the customer's allegation of ¿b30 scope communication error¿ was not confirmed.The device evaluation found the device failed the leak test at the bending cover, the control body had corrosion inside, switches 3 and 4 were not working, the device failed electrical continuity test @ ol ohm.The investigation is ongoing and follow-up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the results of the investigation, it is likely the following led to the malfunction: olympus confirmed the biopsy channel was leaking while the user was handling the device.Due to the water invasion abnormality of electronic parts occurred which led to the malfunction.The event can be detected/prevented by following the instructions for use which state: important information ¿ please read before use precautions: caution turn the video system center on only when the endoscope connector is connected to the light source.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.Precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system inspection of the endoscopic image confirm that the wli and nbi endoscopic images are normal.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|