AIZU OLYMPUS CO., LTD. EVIS EXERA III BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-P190
Device Problems Leak/Splash (1354); Mechanical Problem (1384); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/19/2023
Event Type  malfunction  
Event Description
The customer reported to olympus that the evis exera iii bronchovideoscope had angulation malfunction, the light guide lens had defects, and a b30 scope communication error.There were no reports of patient or user harm associated with this event.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation of ¿b30 scope communication error¿ was not confirmed.The device evaluation found the device failed the leak test at the bending cover, the control body had corrosion inside, switches 3 and 4 were not working, the device failed electrical continuity test @ ol ohm.The investigation is ongoing and follow-up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the results of the investigation, it is likely the following led to the malfunction: olympus confirmed the biopsy channel was leaking while the user was handling the device.Due to the water invasion abnormality of electronic parts occurred which led to the malfunction.The event can be detected/prevented by following the instructions for use which state: important information ¿ please read before use precautions: caution turn the video system center on only when the endoscope connector is connected to the light source.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.Precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system inspection of the endoscopic image confirm that the wli and nbi endoscopic images are normal.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III BRONCHOVIDEOSCOPE
Type of Device
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17933860
MDR Text Key325641418
Report Number9610595-2023-15164
Device Sequence Number1
Product Code EOQ
UDI-Device Identifier04953170342110
UDI-Public04953170342110
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201758
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBF-P190
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-