CARDIAC SCIENCE CORPORATION POWERHEART G5 KIT, AUTO, INTL ENG; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number G5A-00A
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Complainant alleged that during biomed testing, the device displayed a "red x" indicator.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device activity log.However, the device was put through extensive testing including functionality testing without duplicating the report.Internal inspection of the device found no discrepancies.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERHEART G5 KIT, AUTO, INTL ENG
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
CARDIAC SCIENCE CORPORATION
500 burdick parkway
deerfield WI 53531 9692
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key17939906
MDR Text Key326819689
Report Number2112020-2023-00931
Device Sequence Number1
Product Code MJK
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P160033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberG5A-00A
Device Catalogue NumberG5A-00A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-