Model Number 8300AB
Device Problem Perivalvular Leak (1457)
Patient Problems Chest Pain (1776); Dyspnea (1816); Dizziness (2194); Peripheral Edema (4578)
Event Date 09/20/2023
Event Type  Injury  
Event Description
It was reported and learned through medical records that a patient with a 21mm 8300ab aortic valve is under evaluation for a valve in valve procedure after an implant duration of one year, one month due to moderate perivalvular leak.Patient reports ongoing chest pain, shortness of breath, dizziness, and leg edema.The procedure has not been scheduled yet.
 
Manufacturer Narrative
H10: additional manufacturer narrative: the subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Manufacturer Narrative
The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Engineering evaluation summary: per technical summary 33069, rev.A, regurgitation is considered to be a perivalvular leak (pvl) if a turbulent eccentric jet originates between the bioprosthetic sewing ring and the annulus.Pvl can occur in the mitral and aortic position for similar reasons.In the early postoperative period, the highest incidence of pvl has been seen in patients developing infective endocarditis, which is most likely attributed to inadequate peri-operative antibiotic prophylaxis or nosocomial infection.Annular calcification is also a risk factor for the development of peri-operative pvl as the bioprosthesis may not seat properly after debridement.Technique related factors, such as incorrect valve sizing, have been shown to contribute to the development of pvl.Anatomical factors may create difficulty seating the bioprosthetic valve resulting in pvl.The anatomy of the annulus may induce mechanical stresses along the rigid bioprosthetic ring which can influence long-term valve performance and durability.A diseased or rigid annulus can potentially increase the mechanical stress on the prosthetic valve, leading to pvl.Additionally, relevant technical summary 33069, rev.A, exists for this issue.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.A definitive root cause cannot be conclusively determined; however, patient and/or procedural factors likely caused or contributed.All pertinent information available to edwards lifesciences has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key17940068
MDR Text Key325740457
Report Number2015691-2023-16838
Device Sequence Number1
Product Code LWR
UDI-Device Identifier00690103194487
UDI-Public(01)00690103194487(17)250726
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number8300AB
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age61 YR
Patient SexFemale
-
-