Model Number CI-1601-05 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Skin Erosion (2075)
|
Event Date 08/31/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient reportedly experienced breakdown at the incision site.The recipient's device was explanted.The recipient will not be reimplanted.The recipient has reportedly healed.
|
|
Manufacturer Narrative
|
This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|