DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Failure to Select Signal (1582)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) displaying signal difficulties.The rn, called for assistance to help obtain blood pressure on the unit.The console was changed out and the nurse stated" there were so many problems with this one already and that the service people have looked at it several times already." there was no patient harm or injury reported.
 
Manufacturer Narrative
Updated fields: b4, d4, d9, g3, g6, h2, h3, h4, h6(type of investigation, investigation findings, component, investigation conclusions), h10.A getinge field service engineer (fse) was dispatched to investigate the issue.The fse stated that the fiber optic module was previously replaced on the unit because the customer was complaining of no pressure wave form on the screen.The fse then stated that replacing that part did not resolve the issue.So, the fse ordered a front end board.After replacing the front end board the fse performed system checkout.The fse then stated that the unit always showed the pressure waveform with his fiber optic tester and with the trainer.The equipment passed all functional testing.The service order was then closed and the unit was cleared for clinical use.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17956185
MDR Text Key325906249
Report Number2249723-2023-04491
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2009
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-