It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) displaying signal difficulties.The rn, called for assistance to help obtain blood pressure on the unit.The console was changed out and the nurse stated" there were so many problems with this one already and that the service people have looked at it several times already." there was no patient harm or injury reported.
|
Updated fields: b4, d4, d9, g3, g6, h2, h3, h4, h6(type of investigation, investigation findings, component, investigation conclusions), h10.A getinge field service engineer (fse) was dispatched to investigate the issue.The fse stated that the fiber optic module was previously replaced on the unit because the customer was complaining of no pressure wave form on the screen.The fse then stated that replacing that part did not resolve the issue.So, the fse ordered a front end board.After replacing the front end board the fse performed system checkout.The fse then stated that the unit always showed the pressure waveform with his fiber optic tester and with the trainer.The equipment passed all functional testing.The service order was then closed and the unit was cleared for clinical use.
|