OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problems No Audible Alarm (1019); Alarm Not Visible (1022); Material Separation (1562)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Other text: b3: date of event and d4: udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the devices' alarm was not working and the light emitting diode (led) alarm indicators are broken off the board.Patient involvement is unknown.
 
Manufacturer Narrative
Evaluation codes: updated.D3, g1,2 email is: (b)(6) no product was returned, and no photographic evidence was provided to aid in this investigation.As a result, a complaint investigation/product evaluation and problem confirmation could not be performed.The cause could not be determined; however, based on the reported problem, the most probable cause was a damaged light emitting diode (led) indicator from the printed circuit board (pcb) board due to cover misalignment during closure.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.If the product is returned the manufacturer will re-open the complaint for further device analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17957865
MDR Text Key325951226
Report Number3012307300-2023-09719
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/23/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-