• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN PROFIX TOTAL KNEE TIB BASEPLATE ALL-POLY; PROSTHESIS, TOE, CONSTRAINED, POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN PROFIX TOTAL KNEE TIB BASEPLATE ALL-POLY; PROSTHESIS, TOE, CONSTRAINED, POLYMER Back to Search Results
Catalog Number UNKN01200615
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Joint Laxity (4526)
Event Date 03/26/2018
Event Type  Injury  
Manufacturer Narrative
Internal reference number: (b)(4).H10: this complaint was opened by smith+nephew to document a patient complication identified through a review of clinical evidence from literature sources that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.Alexander w.Glover, alasdair j.A.Santini, john s.Davidson, jill a.Pope.Mid- to long-term survivorship of oxidised zirconium total knee replacements performed in patients under 50years of age,the knee, volume 25, issue 4, 2018, pages 617-622, issn 0968-0160, https://doi.Org/10.1016/j.Knee.2018.03.014.
 
Event Description
It was reported that on literature review "mid- to long-term survivorship of oxidised zirconium total knee replacements performed in patients under 50 years of age", 2 patients required tibial revision due to inadequate posterior slope.Both patients improved their flexion post-operatively.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN PROFIX TOTAL KNEE TIB BASEPLATE ALL-POLY
Type of Device
PROSTHESIS, TOE, CONSTRAINED, POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17961677
MDR Text Key326007682
Report Number1020279-2023-01990
Device Sequence Number1
Product Code KWH
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Literature
Reporter Occupation Other
Type of Report Initial
Report Date 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKN01200615
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received10/17/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-