OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400 Back to Search Results
Model Number WB91051W
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to an olympus repair center for evaluation and the customer¿s reported issue was confirmed.The device evaluation found that the e433 error was due to a defective generator board.The e133 error was unable to be reproduced however the e172 was recorded in the fault log and the low voltage power supply board was replaced.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event Description
The customer reported that generator has an e433 (rebooting ) error and a e133 error code.There were no reports of harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF UNIT "ESG-400"
Type of Device
ELECTROSURGICAL GENERATOR ESG-400
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17961905
MDR Text Key326024338
Report Number9610773-2023-02987
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04042761076838
UDI-Public04042761076838
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWB91051W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-