Patient reported that they were unable to calibrate their device.When performing control solution testing, the member received the following results: control solution 1: range - 102-153., results - 70, and 79; control solution 2 range - 275-413, results - 258, and 252.
|
The test strips and control solution associated with this report were not returned for testing.Review of the dhr for both the test strips and control solution used at the time of the event was performed and no non-conformances were noted.Root cause of the patient receiving out of range control solution tests was unable to be determined.The blood glucose meter that was used at the time of the event was returned for investigation, and testing was performed.Testing was able to be completed successfully, with no issues noted.
|