Based on the log file analysis the ventilator sporadically detected an expiratory flow during inspiration (pmax valve err).It can be assumed, that a (temporarily) too slow electronic peep valve (mv1) caused a delayed closure of the pneumatic peep valve.In consequence, the device alarmed for reinstall vent, as reported.The applied volumes and resulting airway pressures were occasionally reduced.No indication for a faulty ventilator motor was given.A sticking/stucking peep valve is detected during automatic power-on-self-test.During automatic ventilation, pressure control might be disturbed, as consequence airway pressures, tidal and minute volumes deviating from user settings are possible.Depending on the set alarm limits, alarms regarding inspiratory pressure and/or tidal/minute volume will be generated, as well as technical alarms.Manual ventilation and monitoring functions are not affected.
|